STANDARD FEATURES
Ranges:
to 1000 psig
Accuracy:
0.1% of full scale
Repeatability: 0.03% of full scale
Sensitivity:
0.01% of full scale
Hysteresis: 0.1% of full scale for ranges of 150 psi and below;
0.2% of full scale for 200 psi and above
Maximum
temperature effect: 0.1% of full scale per 10°C/50°F change from
23°C/73.4°F
Scale length:
30" through two pointer revolutions
Dial
diameter: 6"
Pressure
element volume:
Ranges 150 psi
and below: 6.9 cc with pointer at zero; 8.6 cc, at full scale.
Ranges 200 psi
and above: 10 cc. For the range 1-125 water only: 6.3 cc with pointer at
zero; 8.0 cc, at full scale
Liquid
service: Gauges with ranges of 200, 300, and 500 psi (Bourdon tubes)
can be calibrated for liquid service as an option. The 0-1000 psi range is
calibrated for liquid service only.
Case
pressure: Cases are open to atmosphere.
Bleed tube:
Ranges of 200, 300, and 500 psi (Bourdon tubes), when calibrated for
liquid service, have bleed tubes to facilitate filling and flushing. On
the 1000 psi range, a bleed tube is standard.
Case
connections: 150 psi and below, 1/8" female NPT; 200 psi and above, ¼"
female NPT. Both have a built-in stainless steel filter and are located in
the bottom of the case.
Case
construction: Anodized aluminum with high-strength, scratch-resistant
plastic dial cover. Bezel has no screw holes; the case has special clips,
which make flush mounting easy.
Zero
adjustment: A zero adjustment is accessible through the front of the
case.